வேதியியல் அறிவியல் இதழ்

ஐ.எஸ்.எஸ்.என்: 2150-3494

திறந்த அணுகல்
கையெழுத்துப் பிரதியை சமர்ப்பிக்கவும் arrow_forward arrow_forward ..

Fatigue Test and Lifetime Prediction of Implantable Biocomposite Materials

Abstract

Jin-Young Choi, Jisoo Kim, Jinyong Shin, Minsu Lee,Tae-Ho Yoon and Sunmook Lee

Biocomposite materials consisting of Poly-L-Lactic-Co-Glycolic Acid (PLGA) and Hydroxyapatite (HA) were synthesized and mechanical properties have been observed by carrying out the fatigue test, for which accelerated tests have been conducted such that 20%, 30% and 40% of the yield strength have been applied in a repeated mode until a crack occurred on the load-applied points of test specimen. Based on the fatigue test results, the lifetime of test specimen (B10 life) was estimated. It was found that Weibull distribution (shape parameter=8.79) was the most proper one to describe the failure data obtained from the accelerated fatigue tests. B10 life at 5% of yield strength was estimated to be 335,764 cycles (Confidence Level=95%, lower limit) by applying an inverse power model.

மறுப்பு: இந்த சுருக்கமானது செயற்கை நுண்ணறிவு கருவிகளைப் பயன்படுத்தி மொழிபெயர்க்கப்பட்டது மற்றும் இன்னும் மதிப்பாய்வு செய்யப்படவில்லை அல்லது சரிபார்க்கப்படவில்லை

இந்தக் கட்டுரையைப் பகிரவும்

குறியிடப்பட்டது